Fatork SP506 User Manual

The Fatork SP506 is a high-performance digital pressure calibrator designed for precision measurement and calibration of pressure instruments. It features a high-resolution display, multiple pressure units, data logging, HART communication capability, and rugged construction for field use. Below are key sections for safety, features, installation, setup, operation, and troubleshooting.

1. Features at a Glance 2. Installation 3. First-Time Setup 4. Specifications 5. Controls and Settings 6. Connecting Devices 7. Measurement Operations 8. Data Management 9. Care & Maintenance 10. Regulatory Information 11. Troubleshooting

Features at a Glance

Key components: High-resolution LCD display, Pressure ports (input/output), Function buttons, HART communication port, USB interface, Battery compartment.

FeatureDescription
Pressure RangeDual range: 0-100 psi and 0-1000 psi (depending on model)
Accuracy±0.05% of full scale
DisplayBacklit LCD with digital readout
Unitspsi, bar, kPa, MPa, inHg, mmHg, mH2O
HART CommunicationFor smart transmitter calibration
Data LoggingStores up to 1000 calibration records
BatteryRechargeable lithium-ion (8+ hours operation)
Protection RatingIP54 dust and water resistant
Pressure Ports1/4" NPT female (input and output)
Zero ButtonQuick zero calibration at current atmospheric pressure

Installation

Unpack and inspect contents. Ensure all accessories are present.

  1. Battery installation: Open battery compartment, insert rechargeable battery, secure cover.
  2. Pressure connections: Use appropriate adapters for 1/4" NPT connections.
  3. Initial charging: Connect to AC adapter and charge fully before first use (4 hours).
  4. Protective case: Use provided carrying case for field protection.

WARNING! Do not exceed maximum pressure rating. Use appropriate personal protective equipment when working with pressurized systems.

First-Time Setup

Power on calibrator; follow initial configuration prompts.

  1. Language selection: Press MENU > SETTINGS > LANGUAGE.
  2. Unit selection: Configure preferred pressure units (default: psi).
  3. Date/Time setup: Enter current date and time for accurate data logging.
  4. Zero calibration: Press ZERO button in stable environment to set reference.
  5. Communication setup: Configure HART and USB settings if needed.

CAUTION! Perform zero calibration at stable temperature away from drafts or vibration sources.

Specifications

Controls and Settings

Front Panel: POWER, ZERO, MENU, ENTER, UP/DOWN arrows, ESC, HART button.

Display Indicators: Battery level, Unit, HART status, Data logging, Error codes.

Menu Structure: MEASURE (live pressure reading), CALIBRATE (calibration procedures), LOGGER (data management), SETTINGS (unit, language, date/time, communication), SYSTEM (diagnostics, firmware info).

Calibration Modes: Zero calibration, Span calibration, Full calibration with reference standard.

Connecting Devices

Pressure connections: Use clean, dry pressure source. HART: Connect to HART communicator or compatible transmitter. USB: For PC connection and data transfer.

  1. Ensure pressure system is depressurized before connecting.
  2. Use appropriate seals and thread tape for pressure connections.
  3. For HART communication: Connect between transmitter and loop power supply.
  4. USB connection: Use included cable for PC software interface.

Tip: Always bleed pressure slowly and verify connections are leak-free before applying full pressure.

Measurement Operations

Supports gauge, absolute, and differential pressure measurements.

  1. Power on: Press and hold POWER for 2 seconds.
  2. Select mode: Use MENU to choose MEASURE mode.
  3. Apply pressure: Connect to pressure source, apply slowly to target pressure.
  4. Record reading: Use ENTER to capture measurement or start continuous logging.
  5. Zero check: Press ZERO to recalibrate at atmospheric pressure.
  6. Unit change: Press UNIT to cycle through available pressure units.
  7. HART communication: Press HART button to access smart device functions.
  8. Data save: Press LOG to save current reading with timestamp.

WARNING! Do not subject to pressure spikes or overpressure conditions beyond rated maximum.

Data Management

Data Logging: Stores up to 1000 records with timestamp and measurement data. Data Transfer: USB connection to PC with Fatork Data Manager software. Report Generation: Create calibration certificates and reports via software.

Care & Maintenance

Regular cleaning: Wipe with soft, dry cloth. Pressure port cleaning: Use appropriate cleaners for port threads. Battery maintenance: Charge fully before storage; avoid complete discharge. Calibration interval: Recommended annual calibration by authorized service center.

CAUTION! Do not immerse in liquids. Avoid exposure to corrosive chemicals. Store in protective case when not in use.

Regulatory Information

CE Certified: Complies with European EMC and safety directives. FCC Part 15: Class A digital device; may cause interference in residential areas. RoHS Compliant: Contains no hazardous substances above permitted levels. Manufactured under ISO 9001 quality system. For support, visit www.fatork.com or contact technical support at +1-800-555-3476.

Troubleshooting

SymptomPossible CauseCorrective Action
No displayBattery depletedConnect to charger; allow 30 minutes before use.
Inaccurate readingZero drift, contaminationPerform zero calibration; clean pressure ports.
HART not communicatingConnection, loop powerCheck HART connections; verify 4-20mA loop has sufficient power.
USB not recognizedDriver, cable, portInstall Fatork drivers; try different USB cable/port.
Error code displayedSensor fault, overpressureNote error code; refer to manual appendix; contact support if persistent.
Data logging fullMemory capacity reachedTransfer data to PC; clear memory via LOGGER menu.

Factory Reset: MENU > SYSTEM > RESET TO FACTORY (requires password: 0000).

Service: +1-800-555-3476; www.fatork.com/support